How to Adjust the Pressure of a Respironics CPAP Machine

Table of contents:

How to Adjust the Pressure of a Respironics CPAP Machine
How to Adjust the Pressure of a Respironics CPAP Machine
Anonim

Here are instructions for adjusting the pressure (and other settings) of a Respironics REMStar Plus CPAP machine.

Steps

Adjust Pressure on a Respironics CPAP Machine Step 1
Adjust Pressure on a Respironics CPAP Machine Step 1

Step 1. Find the model you have

Methods for accessing clinical settings vary from machine to machine. Try one of the following:

Adjust Pressure on a Respironics CPAP Machine Step 2
Adjust Pressure on a Respironics CPAP Machine Step 2

Step 2. With the machine connected to the power supply, remove the cable from the instrument

Then hold down the two buttons below the monitor as you plug the cable back into the machine. Keep pressing the buttons until you hear a “beep”. This indicates that the CPAP machine has entered the Therapy Settings Menu.

  • For RESMED S7, S8, and R241 models: hold down the right button for 5 seconds.
  • For the RESMED S6 model: press the Start button and the 20Min button at the same time while the unit is on. The two buttons must remain lit. Now press '5' to lower the pressure and '10' to increase it. Turn off the machine to save the settings.
Adjust Pressure on a Respironics CPAP Machine Step 3
Adjust Pressure on a Respironics CPAP Machine Step 3

Step 3. Reset the display

It will first show you the amount of hours the machine has been used. To cancel this value and start over from scratch, hold down the humidifier or ramp button. An "X" will appear. Keep holding the key until the value becomes “0” and the “X” disappears.

To skip this setting or continue, press the right button

Adjust Pressure on a Respironics CPAP Machine Step 4
Adjust Pressure on a Respironics CPAP Machine Step 4

Step 4. Set the 'quota'

This value will be what appears next. To change it, press the humidifier or ramp button until the desired setting is reached. The altitude values are: 1 = less than 760 meters; 2 = from 760 to 1500 meters; 3 = from 1501 to 2300 meters.

To skip this setting or continue, press the right button

Adjust Pressure on a Respironics CPAP Machine Step 5
Adjust Pressure on a Respironics CPAP Machine Step 5

Step 5. The "therapy" mode will appear:

CPAP or CFLEX.

To change it, press the ramp or humidifier button until the desired value is reached.

To skip this setting or continue, press the right button

Adjust Pressure on a Respironics CPAP Machine Step 6
Adjust Pressure on a Respironics CPAP Machine Step 6

Step 6. CPAP pressure will appear

To change the value, press the ramp or humidifier button until the desired value is reached.

To skip this setting or continue, press the right button

Adjust Pressure on a Respironics CPAP Machine Step 7
Adjust Pressure on a Respironics CPAP Machine Step 7

Step 7. The CPAP adjustment will appear

This will allow you to calibrate the pressure with a pressure gauge. It is suggested that you do not try to change this setting.

To skip this setting or continue, press the right button

Adjust Pressure on a Respironics CPAP Machine Step 8
Adjust Pressure on a Respironics CPAP Machine Step 8

Step 8. The CFLEX setting will appear if you select the CFLEX mode in step number 5

If you choose setting 1 you will have a minimum pressure drop, setting 3 the maximum drop. To change this value, press the ramp or humidifier button until you reach the desired one.

To skip this step or continue, press the right button

Adjust Pressure on a Respironics CPAP Machine Step 9
Adjust Pressure on a Respironics CPAP Machine Step 9

Step 9. The ramp time will now appear

It could be set from 0 to 45 minutes in 5 minute increments. To change these values, press the ramp or humidifier button until you reach the desired values.

To skip this step or continue, press the right button

Adjust Pressure on a Respironics CPAP Machine Step 10
Adjust Pressure on a Respironics CPAP Machine Step 10

Step 10. The initial ramp pressure will appear

To change this setting, press the humidifier or ramp button until the desired value is reached.

To skip this step or continue, press the right button

Adjust Pressure on a Respironics CPAP Machine Step 11
Adjust Pressure on a Respironics CPAP Machine Step 11

Step 11. The patient disconnect setting will now appear

This is an alarm that is activated if there is a leak in the mask and turns off the air flow. 1 = Alarm active; 2 = Alarm off. To change this setting press the ramp or humidifier button.

Adjust Pressure on a Respironics CPAP Machine Step 12
Adjust Pressure on a Respironics CPAP Machine Step 12

Step 12. There are no other settings available. Press the start / stop button to exit the menu, or press the right button and review the settings one more time.